Integrated Table Motion (ITM)
Baxter Hillrom TS1000dv Surgical Table
-
Wireless communication between the table and the da Vinci Xi allows movement of the OR table for patient repositioning without undocking.
-
The only operating table compatible with Integrated Table Motion is the Hillrom TS7000dV table. All other operating tables are incompatible with Integrated Table Motion, including the Hillrom TS7000. Always undock the system from the patient before moving an incompatible table.
-
To reapply the instrument arms brake and be ready for surgical use, disengage ITM once repositioning is done.
-
Users press the DA VINCI (dV) button on the remote control to turn on Integrated Table Motion, adjust the table with the positioning buttons, then press the DA VINCI (dV) button again to turn off Integrated Table Motion.
-
Radio frequency is used for the wireless communication between the table and the da Vinci Xi to ensure coordinated movements.
-
Always check patient clearance, port sites, cannula remote centers, and instrument tips before, during, and after each use of Integrated Table Motion.