Skip to main content

Spinal Drainage System

 Sent on behalf of Philip Lau - April 2, 2024
 
Situation:
· All Medtronic DUET spinal drain systems have been removed due to safety recall. 
Background:
· Inadvertent disconnections of the Medtronic DUET spinal drain system have been reported, leading to CSF leakage and potential infection risk. As such, there has been a world-wide recall of the product since Jan 2024. 
Assessment:
· There are only 2 known alternatives currently on the market. Quantities are also extremely limited due to the world-wide demand – all products are currently on allocation.
1. Medtronic Becker – instructions attached
   §  Already in use as a sub throughout the hospital.
   §  Last month Sunnybrook’s usage ~20 each.
   §  1 disconnection at the drainage bag noted locally, peer hospital  (TWH) noted 5 disconnections.
   §  No current recall notice for this product.
2. Integra Codman EDS3 – instructions attached
   §  Sunnybrook is currently allocated only 1 system per week, totaling 11 systems in hospital.
   §  TWH has already switched exclusively to this product due to concerns over the Medtronic Becker system. 
Recommendation:
· Remove all Medtronic DUET systems (already in place since Feb 2024).
· Switch over to Integra Codman EDS3 to mitigate potential issues with the Medtronic Becker system; however, the hospital does not currently have a sustainable supply.
· Additional in-services for the Integra Codman EDS3 will be provided to all clinical areas this week.
· Medtronic Becker systems will continue to be used once the hospital’s supply of the Integra Codman EDS3 is depleted.
· As the supplies are extremely limited and uncertain, all available stock are being stored centrally in A2-10e, Margaret (Max) Lo Dico’s office. During off hours, call the CrCU Team Leader for access to supplies. There will be a sign-out sheet to log the date, patient name, and location to track usage. During this time, only 1 system at a time will be released as patient need arises.
 
SUPPLY ACCESS
Room: A2-10e (turn left at CrCU entrance doors)
M-F 0800-1600
Max Lo Dico
x88110 or 437-229-6307
 
All other times
CrCU Team Leader
x88116